2015 New Jersey Revised Statutes
Title 34 - LABOR AND WORKMEN'S COMPENSATION
Section 34:15-35.16 - Hearing tests; instruments; test conditions

NJ Rev Stat § 34:15-35.16 (2015) What's This?

34:15-35.16. Hearing tests; instruments; test conditions
Hearing levels shall be determined at all times by using puretone air-conduction audiometric instruments calibrated in accordance with American National Standard ANSI S3.6-1969-R 1973 and ANSI S3.13-1972 and performed in an environment as prescribed by American National Standard S3.1-1960 R 1971 (American Standard Criteria for Background Noise in Audiometer Rooms). To measure permanent hearing loss, hearing tests shall be performed after at least 16 hours absence from exposure to hazardous noise. The calibration of an audiometric instrument used to measure permanent hearing loss shall have been performed within 1 year of the time of the hearing examination, to assure that the audiometer is within the tolerances permitted by the ANSI standards.

L.1979, c. 285, s. 7.


Disclaimer: These codes may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.

This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.