34:15-35.16 - Hearing tests;  instruments;  test conditions

34:15-35.16.  Hearing tests;  instruments;  test conditions
    Hearing levels shall be determined at all times by using puretone air-conduction audiometric instruments calibrated in accordance with American National Standard ANSI S3.6-1969-R 1973 and ANSI S3.13-1972 and performed in an  environment as prescribed by American National Standard S3.1-1960 R 1971 (American Standard Criteria for Background Noise in Audiometer Rooms).  To measure permanent hearing loss, hearing tests shall be performed after at least  16 hours absence from exposure to hazardous noise.  The calibration of an  audiometric instrument used to measure permanent hearing loss shall have been  performed within 1 year of the time of the hearing examination, to assure that  the audiometer is within the tolerances permitted by the ANSI standards.

     L.1979, c. 285, s. 7.
 

Disclaimer: These codes may not be the most recent version. New Jersey may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.